Each item is
standard supplied with a certificate of test which is in accordance with BS EN
60584-1:1996 (ITS90). The e.m.f produced will not deviate from the published
table by an amount greater than the limits defined in BS EN 60584-2:1993.
Reference test points (ITS 90) Gold (1064.18oC) and Palladium in air
(1553.5oC). A temperature(oC) e.m.f (mV) plot with error is provided. In the multipoint calibration we work according to
“Richtlinie fur die Kalibrierung von Thermoelementen an Fixpunkten (in DKD in
Vorbereitung) und nach der Richtlinie
fur
die Kalibrierung von technischen Thermoelementen nach der Vergleichsmethode (DKD)”.
Temperature references are Sn 9206, Zn9205, Al9104 and Ag 93.2.
In this multipoint calibration we provide a
temperature(oC),
e.m.f (mV)
plot with error. The temperatures are traceable to the DKD temperature
references.